Regulations last checked for updates: Oct 17, 2024

Title 10 - Energy last revised: Oct 10, 2024
§ 35.655 - Full-inspection servicing for teletherapy and gamma stereotactic radiosurgery units.

(a) A licensee shall have each teletherapy unit and gamma stereotactic radiosurgery unit fully inspected and serviced during each source replacement to assure proper functioning of the source exposure mechanism and other safety components. The interval between each full-inspection servicing shall not exceed 5 years for each teletherapy unit and shall not exceed 7 years for each gamma stereotactic radiosurgery unit.

(b) This inspection and servicing may only be performed by persons specifically licensed to do so by the Commission or an Agreement State.

(c) A licensee shall keep a record of the inspection and servicing in accordance with § 35.2655.

[67 FR 20370, Apr. 24, 2002, as amended at 83 FR 33110, July 16, 2018]
authority: Atomic Energy Act of 1954, secs. 81, 161, 181, 182, 183, 223, 234, 274 (42 U.S.C. 2111,2201,2231,2232,2233,2273,2282,2021; Energy Reorganization Act of 1974, secs. 201, 206 (42 U.S.C. 5841,5846; 44 U.S.C. 3504 note
source: 67 FR 20370, Apr. 24, 2002, unless otherwise noted.