(a) Applicability. This section applies to all claims filed after January 1, 2005, and pending on or after March 23, 2010.
(b) Invocation. (1) The claimant may invoke the presumption by establishing that—
(i) The miner engaged in coal-mine employment for fifteen years, either in one or more underground coal mines, or in coal mines other than underground mines in conditions substantially similar to those in underground mines, or in any combination thereof; and
(ii) The miner or survivor cannot establish entitlement under § 718.304 by means of chest x-ray evidence; and
(iii) The miner has, or had at the time of his death, a totally disabling respiratory or pulmonary impairment established pursuant to § 718.204, except that § 718.204(d) does not apply.
(2) The conditions in a mine other than an underground mine will be considered “substantially similar” to those in an underground mine if the claimant demonstrates that the miner was regularly exposed to coal-mine dust while working there.
(3) In a claim involving a living miner, a miner's affidavit or testimony, or a spouse's affidavit or testimony, may not be used by itself to establish the existence of a totally disabling respiratory or pulmonary impairment.
(4) In the case of a deceased miner, affidavits (or equivalent sworn testimony) from persons knowledgeable of the miner's physical condition must be considered sufficient to establish total disability due to a respiratory or pulmonary impairment if no medical or other relevant evidence exists which addresses the miner's pulmonary or respiratory condition; however, such a determination must not be based solely upon the affidavits or testimony of any person who would be eligible for benefits (including augmented benefits) if the claim were approved.
(c) Facts presumed. Once invoked, there will be rebuttable presumption—
(1) In a miner's claim, that the miner is totally disabled due to pneumoconiosis, or was totally disabled due to pneumoconiosis at the time of death; or
(2) In a survivor's claim, that the miner's death was due to pneumoconiosis.
(d) Rebuttal—(1) Miner's claim. In a claim filed by a miner, the party opposing entitlement may rebut the presumption by—
(i) Establishing both that the miner does not, or did not, have:
(A) Legal pneumoconiosis as defined in § 718.201(a)(2); and
(B) Clinical pneumoconiosis as defined in § 718.201(a)(1), arising out of coal mine employment (see § 718.203); or
(ii) Establishing that no part of the miner's respiratory or pulmonary total disability was caused by pneumoconiosis as defined in § 718.201.
(2) Survivor's claim. In a claim filed by a survivor, the party opposing entitlement may rebut the presumption by—
(i) Establishing both that the miner did not have:
(A) Legal pneumoconiosis as defined in § 718.201(a)(2); and
(B) Clinical pneumoconiosis as defined in § 718.201(a)(1), arising out of coal mine employment (see § 718.203); or
(ii) Establishing that no part of the miner's death was caused by pneumoconiosis as defined in § 718.201.
(3) The presumption must not be considered rebutted on the basis of evidence demonstrating the existence of a totally disabling obstructive respiratory or pulmonary disease of unknown origin.
[78 FR 59114, Sept. 25, 2013]
The following standards are established in accordance with sections 402(f)(1)(D) and 413(b) of the Act. They were developed in consultation with the National Institute for Occupational Safety and Health (NIOSH) of the Centers for Disease Control and Prevention in the Department of Health and Human Services. These standards are promulgated for the guidance of physicians and medical technicians to ensure that uniform procedures are used in administering and interpreting X-rays and that the best available medical evidence will be submitted in connection with a claim for black lung benefits. If it is established that one or more standards have not been met, the claims adjudicator may consider such fact in determining the evidentiary weight to be assigned to the physician's report of an X-ray.
(a) Definitions. (1) Digital radiography systems, as used in this context, include both digital radiography (DR) and computed radiography (CR). Digital radiography is the term used for digital X-ray image acquisition systems in which the X-ray signals received by the image detector are converted nearly instantaneously to electronic signals without moveable cassettes. Computed radiography is the term for digital X-ray image acquisition systems that detect X-ray signals using a cassette-based photostimulable storage phosphor. Subsequently, the cassette is processed using a stimulating laser beam to convert the latent radiographic image to electronic signals which are then processed and stored so they can be displayed.
(2) Qualified medical physicist means an individual who is trained in evaluating the performance of radiographic equipment including radiation controls and facility quality assurance programs, and has the relevant current certification by a competent U.S. national board, or unrestricted license or approval from a U.S. State or Territory.
(3) Radiographic technique chart means a table that specifies the types of cassette, intensifying screen, film or digital detector, grid, filter, and lists X-ray machine settings (timing, kVp, mA) that enables the radiographer to select the correct settings based on the body habitus or the thickness of the chest tissue.
(4) Radiologic technologist means an individual who has met the requirements for privileges to perform general radiographic procedures and for competence in using the equipment and software employed by the examining facility to obtain chest images as specified by the State or Territory and examining facility in which such services are provided. Optimally, such an individual will have completed a formal training program in radiography leading to a certificate, an associate's degree, or a bachelor's degree and participated in the voluntary initial certification and annual renewal of registration for radiologic technologists offered by the American Registry of Radiologic Technologists.
(5) Soft copy means the image of a coal miner's chest radiograph acquired using a digital radiography system, viewed at the full resolution of the image acquisition system using an electronic medical image display device.
(b) General provisions. (1) Facilities must maintain ongoing licensure and certification under relevant local, State, and Federal laws and regulations for all digital equipment and related processes covered by this Appendix. Radiographic equipment, its use and the facilities (including mobile facilities) in which such equipment is used must conform to applicable State or Territorial and Federal regulations. Where no applicable regulations exist regarding reducing the risk from ionizing radiation exposure in the clinical setting, radiographic equipment, its use and the facilities (including mobile facilities) in which such equipment is used should conform to the recommendations in NCRP Report No. 102, NCRP Report No. 105, and NCRP Report No. 147 (incorporated by reference, see § 718.5).
(2) Chest radiographs of miners must be performed:
(i) By or under the supervision of a physician who makes chest radiographs in the normal course of practice and who has demonstrated ability to make chest radiographs of a quality to best ascertain the presence of pneumoconiosis; or
(ii) By a radiologic technologist.
(3) Miners must be disrobed from the waist up at the time the radiograph is given. The facility must provide a dressing area and for those miners who wish to use one, the facility will provide a clean gown. Facilities must be heated to a comfortable temperature.
(4) Before the miner is advised that the examination is concluded, the radiograph must be processed and inspected and accepted for quality standards by the physician, or if the physician is not available, acceptance may be made by the radiologic technologist. In a case of a substandard radiograph, another must be made immediately.
(c) Chest radiograph specifications—film. (1) Every chest radiograph must be a single posteroanterior projection at full inspiration on a film being no less than 14 by 17 inch film. Additional chest films or views must be obtained if they are necessary for clarification and classification. The film and cassette must be capable of being positioned both vertically and horizontally so that the chest radiograph will include both apices and costophrenic angles. If a miner is too large to permit the above requirements, then a projection with minimum loss of costophrenic angle must be made.
(2) Radiographs must be made with a diagnostic X-ray machine having a rotating anode tube with a maximum of a 2 mm source (focal spot).
(3) Except as provided in paragraph (c)(4) of this appendix, radiographs must be made with units having generators that comply with the following:
(i) Generators of existing radiographic units acquired by the examining facility prior to July 27, 1973, must have a minimum rating of 200 mA at 100 kVp;
(ii) Generators of units acquired subsequent to that date must have a minimum rating of 300 mA at 125 kVp. A generator with a rating of 150 kVp is recommended.
(4) Radiographs made with battery-powered mobile or portable equipment must be made with units having a minimum rating of 100 mA at 110 kVp at 500 Hz, or 200 mA at 110 kVp at 60 Hz.
(5) Capacitor discharge and field emission units may be used.
(6) Radiographs must be given only with equipment having a beam-limiting device that does not cause large unexposed boundaries. The use of such a device must be discernible from an examination of the radiograph.
(7) To ensure high quality chest radiographs:
(i) The maximum exposure time must not exceed 50 milliseconds except that with single phase units with a rating less than 300 mA at 125 kVp and subjects with chests over 28 cm postero-anterior, the exposure may be increased to not more than 100 milliseconds;
(ii) The source or focal spot to film distance must be at least 6 feet.
(iii) Medium-speed film and medium-speed intensifying screens are recommended. However, any film-screen combination, the rated “speed” of which is at least 100 and does not exceed 300, which produces radiographs with spatial resolution, contrast, latitude and quantum mottle similar to those of systems designated as “medium speed” may be employed;
(iv) Film-screen contact must be maintained and verified at 6-month or shorter intervals.
(v) Intensifying screens must be inspected at least once a month and cleaned when necessary by the method recommended by the manufacturer;
(vi) All intensifying screens in a cassette must be of the same type and made by the same manufacturer;
(vii) When using over 90 kV, a suitable grid or other means of reducing scattered radiation must be used;
(viii) The geometry of the radiographic system must ensure that the central axis (ray) of the primary beam is perpendicular to the plane of the film surface and impinges on the center of the film.
(8) Radiographic processing:
(i) Either automatic or manual film processing is acceptable. A constant time-temperature technique must be meticulously employed for manual processing.
(ii) If mineral or other impurities in the processing water introduce difficulty in obtaining a high-quality radiograph, a suitable filter or purification system must be used.
(9) An electric power supply must be used that complies with the voltage, current, and regulation specified by the manufacturer of the machine.
(10) A test object may be required on each radiograph for an objective evaluation of film quality at the discretion of the Department of Labor.
(11) Each radiograph made under this Appendix must be permanently and legibly marked with the name and address of the facility at which it is made, the miner's DOL claim number, the date of the radiograph, and left and right side of the film. No other identifying markings may be recorded on the radiograph.
(d) Chest radiograph specifications—digital radiography systems. (1) Every digital chest radiograph must be a single posteroanterior projection at full inspiration on a digital detector with sensor area being no less than 1505 square centimeters with a minimum width of 35 cm. The imaging plate must have a maximum pixel pitch of 200 µm, with a minimum bit depth of 10. Spatial resolution must be at least 2.5 line pairs per millimeter. The storage phosphor cassette or digital image detector must be positioned either vertically or horizontally so that the image includes the apices and costophrenic angles of both right and left lungs. If the detector cannot include the apices and costophrenic angles of both lungs as described, then the two side-by-side images can be obtained that together include the apices and costophrenic angles of both right and left lungs.
(2) Radiographs must be made with a diagnostic X-ray machine with a maximum actual (not nominal) source (focal spot) of 2 mm, as measured in two orthogonal directions.
(3) Radiographs must be made with units having generators which have a minimum rating of 300 mA at 125 kVp. Exposure kilovoltage must be at least the minimum as recommended by the manufacturer for chest radiography.
(4) An electric power supply must be used that complies with the voltage, current, and regulation specified by the manufacturer of the machine. If the manufacturer or installer of the radiographic equipment recommends equipment for control of electrical power fluctuations, such equipment must be used as recommended.
(5) Radiographs must be obtained only with equipment having a beam-limiting device that does not cause large unexposed boundaries. The beam limiting device must provide rectangular collimation. Electronic post-image acquisition “shutters” available on some CR or DR systems that limit the size of the final image and that simulate collimator limits must not be used. The use and effect of the beam limiting device must be discernible on the resulting image.
(6) Radiographic technique charts must be used that are developed specifically for the X-ray system and detector combinations used, indicating exposure parameters by anatomic measurements.
(7) To ensure high quality chest radiographs:
(i) The maximum exposure time must not exceed 50 milliseconds except for subjects with chests over 28 cm posteroanterior, for whom the exposure time must not exceed 100 milliseconds.
(ii) The distance from source or focal spot to detector must be at least 70 inches (or 180 centimeters if measured in centimeters).
(iii) The exposure setting for chest images must be within the range of 100-300 equivalent exposure speeds and must comply with ACR Practice Guidelines for Diagnostic Reference Levels in Medical X-ray Imaging, Section V—Diagnostic Reference Levels for Imaging with Ionizing Radiation and Section VII-Radiation Safety in Imaging (incorporated by reference, see § 718.5). Radiation exposures should be periodically measured and patient radiation doses estimated by the medical physicist to assure doses are as low as reasonably achievable.
(iv) Digital radiography system performance, including resolution, modulation transfer function (MTF), image signal-to-noise and detective quantum efficiency must be evaluated and judged acceptable by a qualified medical physicist using the specifications in AAPM Report No. 93, pages 1-68 (incorporated by reference, see § 718.5). Image management software and settings for routine chest imaging must be used, including routine amplification of digital detector signal as well as standard image post-processing functions. Image or edge enhancement software functions must not be employed unless they are integral to the digital radiography system (not elective); in such cases, only the minimum image enhancement permitted by the system may be employed.
(v)(A) The image object, transmission and associated data storage, film format, and transmissions of associated information must conform to the following components of the Digital Imaging and Communications in Medicine (DICOM) standard (incorporated by reference, see § 718.5):
(1) DICOM Standard PS 3.3-2011, Annex A—Composite Information Object Definitions, sections: Computed Radiographic Image Information Object Definition; Digital X-Ray Image Information Object Definition; X-Ray Radiation Dose SR Information Object Definition; and Grayscale Softcopy Presentation State Information Object Definition.
(2) DICOM Standard PS 3.4-2011: Annex B—Storage Service Class; Annex N—Softcopy Presentation State Storage SOP Classes; Annex O—Structured Reporting Storage SOP Classes.
(3) DICOM Standard PS 3.10-2011.
(4) DICOM Standard PS 3.11-2011.
(5) DICOM Standard PS 3.12-2011.
(6) DICOM Standard PS 13.14-2011.
(7) DICOM Standard PS 3.16-2011.
(B) Identification of each miner, chest image, facility, date and time of the examination must be encoded within the image information object, according to DICOM Standard PS 3.3-2011, Information Object Definitions, for the DICOM “DX” object. If data compression is performed, it must be lossless. Exposure parameters (kVp, mA, time, beam filtration, scatter reduction, radiation exposure) must be stored in the DX information object.
(C) Exposure parameters as defined in the DICOM Standard PS 3.16-2011 must additionally be provided when such parameters are available from the facility digital image acquisition system or recorded in a written report or electronic file and transmitted to OWCP.
(8) A specific test object may be required on each radiograph for an objective evaluation of image quality at the Department of Labor's discretion.
(9) CR imaging plates must be inspected at least once a month and cleaned when necessary by the method recommended by the manufacturer.
(10) A grid or air gap for reducing scattered radiation must be used; grids must not be used that cause Moiré interference patterns in either horizontal or vertical images.
(11) The geometry of the radiographic system must ensure that the central axis (ray) of the primary beam is perpendicular to the plane of the CR imaging plate or DR detector and is correctly aligned to the grid.
(12) Radiographs must not be made when the environmental temperatures and humidity in the facility are outside the manufacturer's recommended range of the CR and DR equipment to be used.
(13) All interpreters, whenever classifying digitally acquired chest radiographs, must have immediately available for reference a complete set of ILO standard digital chest radiographic images provided for use with the Guidelines for the Use of the ILO International Classification of Radiographs of Pneumoconioses (2011 Revision) (incorporated by reference, see § 718.5). Modification of the appearance of the standard images using software tools is not permitted.
(14) Viewing systems should enable readers to display the coal miner's chest image at the full resolution of the image acquisition system, side-by-side with the selected ILO standard images for comparison.
(i)(A) Image display devices must be flat panel monitors displaying at least 3 MP at 10 bit depth. Image displays and associated graphics cards must meet the calibration and other specifications of the Digital Imaging and Communications in Medicine (DICOM) standard PS 3.14-2011 (incorporated by reference, see § 718.5).
(B) Image displays and associated graphics cards must not deviate by more than 10 percent from the grayscale standard display function (GSDF) when assessed according to the AAPM On-Line Report No. 03, pages 1-146 (incorporated by reference, see § 718.5).
(ii) Display system luminance (maximum and ratio), relative noise, linearity, modulation transfer function (MTF), frequency, and glare should meet or exceed recommendations listed in AAPM On-Line Report No. 03, pages 1-146 (incorporated by reference, see § 718.5). Viewing displays must have a maximum luminance of at least 171 cd/m
2, a ratio of maximum luminance to minimum luminance of at least 250, and a glare ratio greater than 400. The contribution of ambient light reflected from the display surface, after light sources have been minimized, must be included in luminance measurements.
(iii) Displays must be situated so as to minimize front surface glare. Readers must minimize reflected light from ambient sources during the performance of classifications.
(iv) Measurements of the width and length of pleural shadows and the diameter of opacities must be taken using calibrated software measuring tools. If permitted by the viewing software, a record must be made of the presentation state(s), including any noise reduction and edge enhancement or restoration functions that were used in performing the classification, including any annotations and measurements.
(15) Quality control procedures for devices used to display chest images for classification must comply with the recommendations of the American Association of Physicists in Medicine AAPM On-Line Report No. 03, pages 1-146 (incorporated by reference, see § 718.5). If automatic quality assurance systems are used, visual inspection must be performed using one or more test patterns recommended by the medical physicist every 6 months, or more frequently, to check for defects that automatic systems may not detect.
(16) Classification of CR and DR digitally-acquired chest radiographs under this Part must be performed based on the viewing images displayed as soft copies using the viewing workstations specified in this section. Classification of radiographs must not be based on the viewing of hard copy printed transparencies of images that were digitally-acquired.
(17) The classification of chest radiographs based on digitized copies of chest radiographs that were originally acquired using film-screen techniques is not permissible.
[79 FR 21613, Apr. 17, 2014]
The following standards are established in accordance with section 402(f)(1)(D) of the Act. They were developed in consultation with the National Institute for Occupational Safety and Health (NIOSH). These standards are promulgated for the guidance of physicians and medical technicians to insure that uniform procedures are used in administering and interpreting ventilatory function tests and that the best available medical evidence will be submitted in support of a claim for black lung benefits. If it is established that one or more standards have not been met, the claims adjudicator may consider such fact in determining the evidentiary weight to be given to the results of the ventilatory function tests.
(1) Instruments to be used for the administration of pulmonary function tests shall be approved by NIOSH and shall conform to the following criteria:
(i) The instrument shall be accurate within ±50 ml or within ±3 percent of reading, whichever is greater.
(ii) The instrument shall be capable of measuring vital capacity from 0 to 7 liters BTPS.
(iii) The instrument shall have a low inertia and offer low resistance to airflow such that the resistance to airflow at 12 liters per second must be less than 1.5 cm H20/liter/sec.
(iv) The instrument or user of the instrument must have a means of correcting volumes to body temperature saturated with water vapor (BTPS) under conditions of varying ambient spirometer temperatures and barometric pressures.
(v) The instrument used shall provide a tracing of flow versus volume (flow-volume loop) which displays the entire maximum inspiration and the entire maximum forced expiration. The instrument shall, in addition, provide tracings of the volume versus time tracing (spirogram) derived electronically from the flow-volume loop. Tracings are necessary to determine whether maximum inspiratory and expiratory efforts have been obtained during the FVC maneuver. If maximum voluntary ventilation is measured, the tracing shall record the individual breaths volumes versus time.
(vi) The instrument shall be capable of accumulating volume for a minimum of 10 seconds after the onset of exhalation.
(vii) The instrument must be capable of being calibrated in the field with respect to the FEV1. The volume calibration shall be accomplished with a 3 L calibrating syringe and should agree to within 1 percent of a 3 L calibrating volume. The linearity of the instrument must be documented by a record of volume calibrations at three different flow rates of approximately 3 L/6 sec, 3 L/3 sec, and 3 L/sec.
(viii) For measuring maximum voluntary ventilation (MVV) the instrument shall have a response which is flat within ±10 percent up to 4 Hz at flow rates up to 12 liters per second over the volume range.
(ix) The spirogram shall be recorded at a speed of at least 20 mm/sec and a volume excursion of at least 10mm/L. Calculation of the FEVl from the flow-volume loop is not acceptable. Original tracings shall be submitted.
(2) The administration of pulmonary function tests shall conform to the following criteria:
(i) Tests shall not be performed during or soon after an acute respiratory illness.
(ii) For the FEV1 and FVC, use of a nose clip is required. The procedures shall be explained in simple terms to the patient who shall be instructed to loosen any tight clothing and stand in front of the apparatus. The subject may sit, or stand, but care should be taken on repeat testing that the same position be used. Particular attention shall be given to insure that the chin is slightly elevated with the neck slightly extended. The subject shall be instructed to expire completely, momentarily hold his breath, place the mouthpiece in his mouth and close the mouth firmly about the mouthpiece to ensure no air leak. The subject will than make a maximum inspiration from the instrument and when maximum inspiration has been attained, without interruption, blow as hard, fast and completely as possible for at least 7 seconds or until a plateau has been attained in the volume-time curve with no detectable change in the expired volume during the last 2 seconds of maximal expiratory effort. A minimum of three flow-volume loops and derived spirometric tracings shall be carried out. The patient shall be observed throughout the study for compliance with instructions. Inspiration and expiration shall be checked visually for reproducibility. The effort shall be judged unacceptable when the patient:
(A) Has not reached full inspiration preceding the forced expiration; or
(B) Has not used maximal effort during the entire forced expiration; or
(C) Has not continued the expiration for least 7 sec. or until an obvious plateau for at least 2 sec. in the volume-time curve has occurred; or
(D) Has coughed or closed his glottis; or
(E) Has an obstructed mouthpiece or a leak around the mouthpiece (obstruction due to tongue being placed in front of mouthpiece, false teeth falling in front of mouthpiece, etc.); or
(F) Has an unsatisfactory start of expiration, one characterized by excessive hesitation (or false starts). Peak flow should be attained at the start of expiration and the volume-time tracing (spirogram) should have a smooth contour revealing gradually decreasing flow throughout expiration; or
(G) Has an excessive variability between the three acceptable curves. The variation between the two largest FEV1's of the three acceptable tracings should not exceed 5 percent of the largest FEV1 or 100 ml, whichever is greater. As individuals with obstructive disease or rapid decline in lung function will be less likely to achieve this degree of reproducibility, tests not meeting this criterion may still be submitted for consideration in support of a claim for black lung benefits. Failure to meet this standard should be clearly noted in the test report by the physician conducting or reviewing the test.
(iii) For the MVV, the subject shall be instructed before beginning the test that he or she will be asked to breathe as deeply and as rapidly as possible for approximately 15 seconds. The test shall be performed with the subject in the standing position, if possible. Care shall be taken on repeat testing that the same position be used. The subject shall breathe normally into the mouthpiece of the apparatus for 10 to 15 seconds to become accustomed to the system. The subject shall then be instructed to breathe as deeply and as rapidly as possible, and shall be continually encouraged during the remainder of the maneuver. Subject shall continue the maneuver for 15 seconds. At least 5 minutes of rest shall be allowed between maneuvers. At least three MVV's shall be carried out. (But see § 718.103(b).) During the maneuvers the patient shall be observed for compliance with instructions. The effort shall be judged unacceptable when the patient:
(A) Has not maintained consistent effort for at least 12 to 15 seconds; or
(B) Has coughed or closed his glottis; or
(C) Has an obstructed mouthpiece or a leak around the mouthpiece (obstruction due to tongue being placed in front of mouthpiece, false teeth falling in front of mouthpiece, etc.); or
(D) Has an excessive variability between the three acceptable curves. The variation between the two largest MVVs of the three satisfactory tracings shall not exceed 10 percent.
(iv) A calibration check shall be performed on the instrument each day before use, using a volume source of at least three liters, accurate to within ±1 percent of full scale. The volume calibration shall be performed in accordance with the method described in paragraph (1)(vii) of this Appendix. Accuracy of the time measurement used in determining the FEV1 shall be checked using the manufacturer's stated procedure and shall be within ±3 percent of actual. The procedure described in the Appendix shall be performed as well as any other procedures suggested by the manufacturer of the spirometer being used.
(v)(A) The first step in evaluating a spirogram for the FVC and FEV1 shall be to determine whether or not the patient has performed the test properly or as described in (2)(ii) of this Appendix. The largest recorded FVC and FEV1, corrected to BTPS, shall be used in the analysis.
(B) Only MVV maneuvers which demonstrate consistent effort for at least 12 seconds shall be considered acceptable. The largest accumulated volume for a 12 second period corrected to BTPS and multiplied by five or the largest accumulated volume for a 15 second period corrected to BTPS and multiplied by four is to be reported as the MVV.