§ 35.657 - Therapy-related computer systems.

The licensee shall perform acceptance testing on the treatment planning system of therapy-related computer systems in accordance with published protocols accepted by nationally recognized bodies. At a minimum, the acceptance testing must include, as applicable, verification of:

(a) The source-specific input parameters required by the dose calculation algorithm;

(b) The accuracy of dose, dwell time, and treatment time calculations at representative points;

(c) The accuracy of isodose plots and graphic displays;

(d) The accuracy of the software used to determine sealed source positions from radiographic images; and

(e) The accuracy of electronic transfer of the treatment delivery parameters to the treatment delivery unit from the treatment planning system.

authority: Atomic Energy Act of 1954, secs. 81, 161, 181, 182, 183, 223, 234, 274 (42 U.S.C. 2111,2201,2231,2232,2233,2273,2282,2021; Energy Reorganization Act of 1974, secs. 201, 206 (42 U.S.C. 5841,5846; 44 U.S.C. 3504 note
source: 67 FR 20370, Apr. 24, 2002, unless otherwise noted.
cite as: 10 CFR 35.657