Regulations last checked for updates: Jan 19, 2025

Title 47 - Telecommunication last revised: Jan 15, 2025
§ 95.2585 - MedRadio RF exposure evaluation.

A MedRadio medical implant device or medical body-worn transmitter is subject to the radiofrequency radiation exposure requirements specified in §§ 1.1307(b) and 2.1093 of this chapter, as appropriate. Applications for equipment authorization of devices operating under this section must demonstrate compliance with these requirements using either computational modeling or laboratory measurement techniques. Where a showing is based on computational modeling, the Commission retains the discretion to request that supporting documentation and/or specific absorption rate (SAR) measurement data be submitted, as described in § 2.1093(d)(1) of this chapter.

[85 FR 18151, Apr. 1, 2020]
authority: 47 U.S.C. 154,303,307
source: 82 FR 41104, Aug. 29, 2017, unless otherwise noted.
cite as: 47 CFR 95.2585